Laboratory Research Grade Two Channel Benchtop pH/mV/ISE Meter Kit
$1,434.00The HI5222K is a kit version of the HI5222 advanced research grade dual channel benchtop pH/ISE/mV meter. The kit version includes additional calibration and maintenance solutions, a spare laboratory pH electrode, and a magnetic stirrer.
- 500 mL each of pH 4.01, 7.01 and 10.01 calibration buffers, cleaning, and storage solution
- HI190M Magnetic stirrer
- Extra HI1131B pH electrode
Thank you for your interest in our products. A representative will contact you within the next business day.
The HI5222K includes the HI5222 advanced research grade dual channel benchtop pH/ISE/mV meter that is completely customizable with a large color LCD, capacitive touch keys, and USB port for computer connectivity. The HI5222 is rich in features including 5 point calibration, selectable resolution, data logging, alarm limits, comprehensive GLP, and much more. It retains simplicity with both dedicated key for routine operation and virtual keys that guide the user through setup options. The HI5222 ensures confidence in measurements with the exclusive Hanna Instruments CAL Check™ feature that alerts the user to potential problems during calibration including if the buffer is contaminated or the probe needs to be cleaned.
The use of a magnetic stirrer allows for proper mixing and the sample to be homogenous in consistency and temperature. The stirring of the sample also helps speed up the response time of a sensor in achieving a stable reading. The HI5222K pH meter (kit) includes the HI190M-1 compact magnetic mini-stirrer
The HI5222K includes two HI1131B glass body, double junction, refillable pH electrodes. The laboratory pH electrodes can be used with samples that range from pH 0-13 and at temperatures between 0 and 100 oC.
The HI5222K includes 500 mL bottle each of pH 4.01, 7.01, and 10.01 pH buffer. All pH calibration buffers are produced in accordance with ISO 3696/BS3978 standards using high purity salts, deionized water, certified weight-checked balances, and Class A glassware in a temperature controlled environment monitored with certified thermometers. Reported values are accurate to +/- 0.01 pH @ 25°C and are traceable to NIST Standard Reference Materials (SRMs).
For maintenance the HI5222K is supplied with HI7061L general purpose cleaning solution and the HI70300L electrode storage solution . When the “clean electrode” message is displayed during the calibration of the HI5222 or there is a poor probe condition (i.e. less than 30%) then soak the pH electrode in the cleaning solution for 15 minutes to clean any deposits from the glass surface. After cleaning the probe should be rinsed with purified water and then allowed to soak in the storage solution for 2-3 hours. The electrode storage solution is also used when the electrode is not in use. Either place a small amount into the storage cap or into a beaker in order to maintain a properly hydrated electrode.
Importance of maintaining a clean pH electrode
The most common cause for pH measurement inaccuracies is an unclean or improperly cleaned electrode. This is very important to note, because during calibration, the instrument assumes that the electrode is clean and that the standardization curve created during the calibration process will remain a valid reference until the next calibration. pH meters on the market today will allow an offset voltage of approximately ±60 mV. The deviation from 0 mV is not unusual but ideally should be no greater than ±30 mV. The calibration process compensates for the change in offset voltage. If the mV offset continues to deviate and the meter is calibrated with a dirty electrode then inaccurate readings will result.
The Cleaning Solution Series ensures maximum efficiency and accuracy of your sensors when used for its designated application. Cleaning is a fast and effective routine that should be performed on a regular basis as a preventative measure against using a dirty electrode and to ensure that the junction is not clogged.
The HI5222K dual channel pH/mV/ISE kit includes the following. Further information and specifications can be found on the pages linked below.
HI5222 is supplied with HI1131B pH electrode, HI7662-T temperature probe, HI76404W electrode holder, HI70004 pH 4.01 buffer solution sachet, HI70007 pH 7.01 buffer solution sachet, HI700601 electrode cleaning solution sachet (2), HI7082 3.5M KCl electrolyte solution (30 mL), 12 VDC adapter and instructions. |
|
115 VAC compact magnetic mini stirrer with Speedsafe™ technology with micro-stir bar (1) |
|
HI1131B double junction, glass body, refillable pH electrode |
|
pH 4.01 calibration buffer (500 mL) |
|
pH 7.01 calibration buffer (500 mL) |
|
pH 10.01 calibration buffer (500 mL) |
|
general purpose cleaning solution (500 ml) |
|
electrode storage solution (500 ml) |
Resources
-
Specifications
Specification Name Detail SKU HI5222K pH Range pH: -2.0 to 20.0, -2.00 to 20.00, -2.000 to 20.000; mV: ±2000 pH Resolution pH: 0.1, 0.01, 0.001; mV: 0.1 pH Accuracy (@25°C/77°F) ±0.1 pH, ±0.01 pH, ±0.002 pH ±1 LSD; ±0.1 mV pH Calibration automatic, up to five point calibration, eight standard buffers available (1.68, 3.00, 4.01, 6.86, 7.01, 9.18, 10.01,12.45) and five custom buffers pH Temperature Compensation automatic or manual from -20.0 to 120.0 °C Temperature Range -20.0 to 120.0 ºC, -4.0 to 248.0 °F, 253.15 to 393.15 K* Temperature Resolution 0.1 °C, 0.1 °F, 0.1 K Temperature Accuracy ±0.2 °C, ±0.4 °F, ±0.2 K ISE Range 1 x 10⁻⁶ to 9.99 x 10¹⁰ concentration ISE Resolution 1; 0.1; 0.01; 0.001 concentration ISE Accuracy @ 25 °C/77 °F ±0.5% (monovalent ions); ±1% (divalent ions) ISE Calibration Points automatic, up to five point calibration, seven fixed standard solutions available (0.01, 0.1, 1, 10, 100, 1000, 10000 choice of concentration) and five custom solutions pH Electrode HI1131B glass body pH electrode with BNC connector and 1 m (3.3’) cable (included) Temperature Probe HI7662-T stainless steel temperature probe with 1 m (3.3’) cable (included) Power Supply 12 VDC adapter (included) Environment 0 to 50°C (32 to 122°F; 273 to 323 K), RH max 95% non-condensing Dimensions 160 x 231 x 94 mm (6.3 x 9.1 x 3.7”) Weight 1.2 kg (2.64 lbs.) Ordering Information HI5222K (115V) is supplies with the HI1131B pH electrode, HI7662-T temperature probe, HI76404W electrode holder, HI70004 pH 4.01 buffer solution sachet, HI70007 pH 7.01 buffer solution sachet, HI700601 electrode cleaning solution sachet (2), HI7082 3.5M KCl electrolyte solution (30 mL), 12 VDC adapter and instructions. HI190M 115 VAC compact magnetic mini stirrer with Speedsafe™ technology with micro-stir bar (1), HI7004C pH 4.01 calibration buffer (500 mL), HI7007C pH 7.01 calibration buffer (500 mL), HI7010C pH 10.01 calibration buffer (500 mL), HI7061L general purpose cleaning solution (500 ml), HI70300L electrode storage solution (500 ml) Special Notes *Reduced to actual probe limits Warranty 2 years