| HI2630 and HI2631 |
using the four-ring conductivity probe
|
| EC Range |
0.00 to 29.99 μS/cm 30.0 to 299.9 μS/cm 300 to 2999 μS/cm 3.00 to 29.99 mS/cm 30.0 to 200.0 mS/cm up to 500.0 mS/cm absolute conductivity ‡
|
| EC Resolution |
0.01 μS/cm 0.1 μS/cm 1 μS/cm 0.01 mS/cm 0.1 mS/cm
|
| EC Accuracy at 25 °C / 77 °F |
±1% of reading (±0.05 μS/cm or 1 digit,whichever is greater)
|
| TDS Range |
0.00 to 14.99 ppm (mg/L) 15.0 to 149.9 ppm (mg/L)150 to 1499 ppm (mg/L) 1.50 to 14.99 g/L 15.0 to 100.0 g/L up to 400.0 g/L (with 0.80 factor) absoluteTDS***
|
| TDS Resolution |
0.01 ppm 0.1 ppm 1 ppm 0.01 g/L
|
| TDS Accuracy at 25 °C / 77 °F |
±1% of reading (±0.03 ppm or 1 digit, whichever is greater
|
| Salinity* Range |
0.0 to 400.0 % NaCl 2.00 to 42.00 PSU 0.00 to 80.00 g/L
|
| Salinity* Resolution |
0.1% NaCl 0.01 PSU 0.01 g/L
|
| Salinity* Accuracy at 25 °C / 77 °F |
±1% of reading
|
| Temperature Range |
−20.0 to 120.0 °C**(−4.0 to 248.0 °F)
|
| Temperature Resolution |
0.1 °C (0.1 °F)
|
| Temperature Accuracy at 25 °C / 77 °F |
±0.5 °C (0.9 °F)
|
| EC calibration |
"Cell factor calibration Six standards available: 84 µS/cm, 1413 μS/cm 5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm, 111.8 mS/cm One point offset: 0.00 μS/cm"
|
| Salinity calibration* (% NaCl only) |
Single point with HI7037 standard
|
| Temperature compensation |
• Automatic, −5 to 100 °C (23 to 212 °F) • NoTC, can be selected to measure absolute conductivity
|
| Conductivity temperature coefficient* |
0.00 to 6.00 %/°C (EC and TDS only) (1.90 %/°C default)
|
| TDS factor* |
0.40 to 0.80 (0.50 default)
|
| DO Salinity compensation |
0 to 40 g/L
|
| Logging (HI2631 only) |
Up to 1000* (400) records organized in: • Manual log-on-demand, maximum 200 logs • Manual log-on-stability, maximum 200 logs • Interval logging, maximum 600 logs (100 lots)*
|
| Notes |
*When operating in Standard Mode in HI2631 only!br>** Temperature limits will be reduced to actual probe limits. *** Absolute conductivity (TDS) is the conductivity value without temperature compensation.
|
| PC interface (HI2631 only) |
USB-C
|
| Power supply |
USB Type C (5 VDC; 500 mA)
|
| Environment |
0 to 50 °C (32 to 122 °F) Maximum 95 % RH non-condensing
|
| Dimensions |
205 x 160 x 77 mm (8.0 x 6.2 x 3.0 inches)
|
| Weight |
Approximately 0.85 kg (1.87 lbs.)
|