HI83326 is a compact, multiparameter photometer for use in pool and spa applications. The meter is one of the most advanced photometers available with an innovative optical design that utilizes a reference detector and focusing lens to eliminate errors from changes in the light source and from imperfections in the glass cuvette. This meter has 12 different programmed methods measuring 11 key water quality parameters and also offers an absorbance measurement mode for performance verification and for users that would like to develop their own concentration versus absorbance curves. Pool and spa parameters include alkalinity, free chlorine, nitrite, and phosphate which are critical to maintaining a healthy system. To save valuable laboratory benchtop space, the HI83326 doubles as a professional pH meter with its digital pH/temperature electrode input. Now one meter can be used for both photometric and pH measurements.
Please note, reagents are not included with this meter but can be purchased here.
Parameter | Range | Resolution | Accuracy (at 25 °C) | Method | Wavelength | Reagent Code |
Absorbance | 0.000 to 4.000 Abs | 0.001 Abs | +/-0.003Abs @ 1.000 Abs | |||
Alkalinity | 0 to 500 mg/L (as CaCO3 ) | 1 mg/L | ±5 mg/L ±5% of reading | Colorimetric Method | 610 nm | HI775-26 |
Bromine | 0.00 to 8.00 mg/L (as Br2 ) | 0.01 mg/L | ±0.08 mg/L ±3% of reading | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, DPD method | 525 nm | HI93716-01 |
Chlorine, Free | 0.00 to 5.00 mg/L (as Cl2 ) | 0.01 mg/L | ±0.03 mg/L ±3% of reading | Adaptation of the EPA DPD method 330.5 | 525 nm | HI93701-01 |
Chlorine, Total | 0.00 to 5.00 mg/L (as Cl2 ) | 0.01 mg/L | ±0.03 mg/L ±3% of reading | Adaptation of the EPA DPD method 330.5 | 525 nm | HI93711-01 |
Copper, High Range | 0.00 to 5.00 mg/L (as Cu) | 0.01 mg/L | ±0.02 mg/L ±4% of reading | Adaptation of the EPA method | 575 nm | HI93702-01 |
Cyanuric Acid | 0 to 80 mg/L (as CYA) | 1 mg/L | ±1 mg/L ±15% of reading | Adaptation of the turbidimetric method | 525 nm | HI93722-01 |
Hardness, Calcium | 0.00 to 2.70 mg/L (as CaCO3 ) | 0.01 mg/L | ±0.11 mg/L ±5% of reading | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Calmagite method | 525 nm | HI93720-01 |
Iron, Low Range | 0.000 to 1.600 mg/L (as Fe) | 0.001 mg/L | ±0.010 mg/L ±8% of reading | Adaptation of the TPTZ Method | 575 nm | HI93746-01 |
Iron, High Range | 0.00 to 5.00 mg/L (as Fe) | 0.01 mg/L | ±0.04 mg/L ±2% of reading | Adaptation of the EPA Phenanthroline method 315B, for natural and treated waters | 525 nm | HI93721-01 |
Nitrate | 0.0 to 30.0 mg/L (as NO3 - N) | 0.1 mg/L | ±0.5 mg/L ±10% of reading | Adaptation of the cadmium reduction method | 525 nm | HI93728-01 |
Ozone | 0.00 to 2.00 mg/L (as O3 ) | 0.01 mg/L | ±0.02 mg/L ±3% of reading | Colorimetric DPD Method | 525 nm | HI93757-01 |
pH | 6.5 to 8.5 pH | 0.1 pH | ±0.1 pH | Adaptation of the Phenol Red method | 525 nm | HI93710-01 |
Specification | Detail |
---|---|
sku | HI83326-01 |
Wavelength Accuracy | ±1 nm |
Light Source | 5 LEDs with 420 nm, 466 nm, 525 nm, 575 nm, and 610 nm narrow band interference filters |
Spectral bandwidth | 8 nm |
Input Channels | 1 pH electrode input and 5 photometer wavelengths |
pH Electrode | digital pH electrode (not included) |
Logging Type | log on demand with user name and sample ID optional input |
Logging Memory | 1000 readings |
Connectivity | USB-A host for flash drive micro-USB-B for power and computer connectivity |
GLP | calibration data for connected pH electrode |
Display | 128 x 64 pixel LCD with backlight |
Battery Type / Life | 3.7 VDC Li-polymer rechargeable battery / >500 photometric measurements or 50 hours of continuous pH measurement |
Power Supply | 5 VDC USB 2.0 power adapter with USB-A to micro-USB-B cable (included) |
Environment | 0 to 50.0oC (32 to 122.0oF) 0 to 95% RH, non-condensing |
Dimensions | 206 x 177 x 97 mm (8.1 x 7.0 x 3.8) |
Weight | 1.0 kg (2.2 lbs.) |
Photometer/Colorimeter Light Detector | silicon photodetector |
Cuvette Type | round, 24.6 mm |
Number of Methods | 128 max. |
Ordering Information | Supplied with sample cuvettes and caps (4 ea.), cloth for wiping cuvettes, USB to micro USB cable connector, power adapter and instruction manual. |
Warranty | 2 years |
Note | *Reagents are not included. |