HI83303 is a compact, multiparameter photometer for use in aquaculture applications. The meter is one of the most advanced photometers available with an innovative optical design that utilizes a reference detector and focusing lens to eliminate errors from changes in the light source and from imperfections in the glass cuvette. This meter has 20 different programmed methods measuring 12 key water quality parameters and also offers an absorbance measurement mode for performance verification and for users that would like to develop their own concentration versus absorbance curves. Aquaculture parameters include alkalinity, calcium, nitrite, and phosphate which are critical to maintaining a healthy system. It also contains parameters specific to either a marine or freshwater environment. To save valuable laboratory benchtop space, the HI83303 doubles as a professional pH meter with its digital pH/temperature electrode input. Now one meter can be used for both photometric and pH measurements.
Parameter | Range | Resolution | Accuracy(at 25 °C) | Method | ||
Absorbance | 0.000 to 4.000 Abs | 0.001 Abs | +/-0.003Abs @ 1.000 Abs | |||
Alkalinity | 0 to 500 mg/L (as CaCO3 ) | 1 mg/L | ±5 mg/L ±5% of reading | Colorimetric Method | 610 nm | HI775-26 |
Alkalinity, Marine | 0 to 300 mg/L (as CaCO3 ) | 1 mg/L | ±5 mg/L ±5% of reading | Colorimetric Method | 610 nm | HI755-26 |
Ammonia, Low Range | 0.00 to 3.00 mg/L (as NH3 -N) | 0.01 mg/L | ±0.04 mg/L ±4% of reading | Adaptation of the ASTM Manual of Water and Environmental Technology, D1426 Nessler method. | 420 nm | HI93700-01 |
Ammonia, Medium Range | 0.00 to10.00 mg/L (as NH3-N) | 0.01 mg/L | ±0.05 mg/L ±5% of reading | Adaptation of the ASTM Manual of Water and Environmental Technology, D1426, Nessler method | 420 nm | HI93715-01 |
Ammonia, High Range | 0.0 to 100.0 mg/L (as NH3 -N) | 0.1 mg/L | ±0.5 mg/L ±5% of reading | Adaptation of the ASTM Manual of Water and Environmental Technology, D1426, Nessler method | 420 nm | HI93733-01 |
Calcium | 0 to 400 mg/L (as Ca2+) | 1 mg/L | ±10 mg/L ±5% of reading | Adaptation of the Oxalate method | 466nm | HI937521-01 |
Calcium, Marine | 200 to 600 mg/L (as Ca2+) | 1 mg/L | ±6% of reading | Adaptation of the Zincon method | 610 nm | HI758-26 |
Chlorine, Free | 0.00 to 5.00 mg/L (as Cl2 ) | 0.01 mg/L | ±0.03 mg/L ±3% of reading | Adaptation of the EPA DPD method 330.5 | 525 nm | HI93701-01 |
Chlorine, Total | 0.00 to 5.00 mg/L (as Cl2 ) | 0.01 mg/L | ±0.03 mg/L ±3% of reading | Adaptation of the EPA DPD method 330.5 | 525 nm | HI93711-01 |
Copper, Low Range | 0.000 to 1.500 mg/L (as Cu) | 0.001 mg/L | ±0.010 mg/L ±5% of reading | Adaptation of the EPA method | 575 nm | HI95747-01 |
Copper, High Range | 0.00 to 5.00 mg/L (as Cu) | 0.01 mg/L | ±0.02 mg/L ±4% of reading | Adaptation of the EPA method | 575 nm | HI93702-01 |
Nitrate | 0.0 to 30.0 mg/L (as NO3 - N) | 0.1 mg/L | ±0.5 mg/L ±10% of reading | Adaptation of the cadmium reduction method | 525 nm | HI93728-01 |
Nitrite, Marine Ultra Low Range | 0 to 200 µg/L (as N02 -N) | 1 µg/L | ±10 µg/L ±4% of reading | Adaptation of the EPA Diazotization method 354.1 | 466 nm | HI764-25 |
Nitrite, Low Range | 0 to 600 µg/L (as NO2 -N) | 1 µg/L | ±20 µg/L ±4% of reading | Adaptation of the EPA Diazotization method 354.1 | 466 nm | HI93707-01 |
Nitrite, High Range | 0 to 150 mg/L (as NO2-) | 1 mg/L | ±4 mg/L ±4% of reading | Adaptation of the Ferrous Sulfate method | 575 nm | HI93708-01 |
Oxygen, Dissolved | 0.0 to 10.0 mg/L (as O2 ) | 0.1 mg/L | ±0.4 mg/L ±3% of reading | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Azide modified Winkler method | 420 nm | HI93732-01 |
pH | 6.5 to 8.5 pH | 0.1 pH | ±0.1 pH | Adaptation of the Phenol Red method | 525 nm | HI93710-01 |
Phosphate, Marine Ultr Low Range | 0 to 200 µg/L (as P) | 1 µg/L | ±5 µg/L ±5% of reading | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 20th edition, Ascorbic Acid method | 610 nm | HI736-25 |
Phosphate, Low Range | 0.00 to 2.50 mg/L (as PO4 3-) | 0.01 mg/L | ±0.04 mg/L ±4% of reading | Adaptation of the Ascorbic Acid method | 610 nm | HI93713-01 |
Phosphate, High Range | 0.0 to 30.0 mg/L(as PO43-) | 0.1 mg/L | ±1.0 mg/L ±4% of reading | Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Amino Acid method | 525 nm | HI93717-01 |
Specification | Detail |
---|---|
sku | HI83303-01 |
Wavelength Accuracy | ±1 nm |
Light Source | 5 LEDs with 420 nm, 466 nm, 525 nm, 575 nm, and 610 nm narrow band interference filters |
Spectral bandwidth | 8 nm |
Input Channels | 1 pH electrode input and 5 photometer wavelengths |
pH Electrode | digital pH electrode (not included) |
Logging Type | log on demand with user name and sample ID optional input |
Logging Memory | 1000 readings |
Connectivity | USB-A host for flash drive micro-USB-B for power and computer connectivity |
GLP | calibration data for connected pH electrode |
Display | 128 x 64 pixel LCD with backlight |
Battery Type / Life | 3.7 VDC Li-polymer rechargeable battery / >500 photometric measurements or 50 hours of continuous pH measurement |
Power Supply | 5 VDC USB 2.0 power adapter with USB-A to micro-USB-B cable (included) |
Environment | 0 to 50.0oC (32 to 122.0oF) 0 to 95% RH, non-condensing |
Dimensions | 206 x 177 x 97 mm (8.1 x 7.0 x 3.8) |
Weight | 1.0 kg (2.2 lbs.) |
Photometer/Colorimeter Light Detector | silicon photodetector |
Cuvette Type | round, 24.6 mm |
Number of Methods | 128 max. |
Ordering Information | Supplied with sample cuvettes and caps (4 ea.), cloth for wiping cuvettes, USB to micro USB cable connector, power adapter and instruction manual. |
Warranty | 2 years |
Note | *Reagents are not included. |