Seamless design. Exceptional performance. Quality measurements. Our Benchtop Meter (HI2600) proves that simplicity is best, combining functionality with precision. Built for regular, everyday testing, our Benchtop Meter is easy to incorporate into your lab routine.
Our Simple Benchtop Meter (HI2600) measures pH, conductivity, or dissolved oxygen through its smart Digital Electrodes. Supplied with a pH Digital Electrode (HI11310). Users can expand to measure more later with the following compatible Hanna Instruments® Digital Electrodes for ORP (HI36180), Conductivity (HI763100), and Dissolved Oxygen (HI764080).
Advantages of our Digital Electrodes include:
| • HI2600-01 pH, EC, and DO Benchtop Meter (115V) |
| • HI11310 pH digital electrode |
| • HI764026 electrode holder with the following accessories: |
| - base plate with integrated pivot pin |
| - cable holder clip, attache |
| • pH calibration kit: |
| - pH 4 buffer solution (2 sachets) |
| - pH 7 buffer solution (4 sachets) |
| - pH 10 buffer solution sachets (2 sachets) |
| • Electrode cleaning solution (2 sachets) |
| • HI920018 1.5 m USB-C to USB-C cable |
| • USB-C power adapter |
| • Instrument quality certificate |
| • Probe / electrode quality certificate |
| • Battery safety insert |
| • Quick reference guide with QR code for instruction manual download |
| Specification | Detail |
|---|---|
| HI2600 | using pH electrode HI11310 |
| pH Range | −2.00 to 16.00 pH −2.000 to 16.000 pH * |
| pH Resolution | 0.01 pH 0.001 pH * |
| pH Accuracy at 25 °C / 77 °F | ±0.01 pH ±0.002 pH * |
| mV Range | ±1000.0 mV |
| mV Resolution | 0.1 mV |
| mV Accuracy at 25 °C / 77 °F | ±0.2 mV |
| Temperature Range (HI11310) | −20.0 to 120.0 °C ** (−4.0 to 248.0 °F) |
| Temperature Resolution (HI11310) | 0.1 °C; (0.1 °F) |
| Temperature Accuracy at 25 °C / 77 °F (HI11310) | ±0.5 °C; (0.9 °F) |
| pH calibration | Automatic, up to three (five*) points Option to select from five (seven*) standard buffers: pH 1.68*, 4.01 or 3.00, 6.86, 7.01, 9.18, 10.01, 12.45* and two custom buffers* |
| HI2600 - using ORP electrode | HI36180 |
| ORP Range | ± 2000.0 mV |
| ORP Resolution | 0.1 mV |
| ORP Accuracy at 25 °C / 77 °F | ±0.2 mV (±999.9 mV) ±1 mV (±2000 mV) |
| Temperature Range (HI36180) | −20.0 to 120.0 °C**(−4.0 to 248.0 °F) |
| Temperature Resolution (HI36180) | 0.1 °C (0.1 °F) |
| Temperature Accuracy at 25 °C / 77 °F (HI36180) | ±0.5 °C (0.9 °F) |
| Calibration | Single point offset |
| Temperature Compensation | Automatic, using integral temperature sensor −5.0 to 100.0 °C (23.0 to 212.0 °F) |
| Logging | Up to 1000* records organized in: • Manual log-on-demand, maximum 200 logs • Manual log-on-stability, maximum 200 logs • Interval logging* |
| HI2600 - using the four-ring conductivity probe | HI763100 |
| EC Range | 0.00 to 29.99 μS/cm 30.0 to 299.9 μS/cm 300 to 2999 μS/cm 3.00 to 29.99 mS/cm 30.0 to 200.0 mS/cm up to 500.0 mS/cm absolute conductivity*** |
| EC Resolution | 0.01 μS/cm 0.1 μS/cm 1 μS/cm 0.01 mS/cm 0.1 mS/cm |
| EC Accuracy at 25 °C / 77 °F | ±1% of reading (±0.05 μS/cm or 1 digit,whichever is greater) |
| TDS Range | 0.00 to 14.99 ppm (mg/L) 15.0 to 149.9 ppm (mg/L)150 to 1499 ppm (mg/L) 1.50 to 14.99 g/L 15.0 to 100.0 g/L up to 400.0 g/L absoluteTDS ‡ (with 0.80 factor) |
| TDS Resolution | 0.01 ppm 0.1 ppm 1 ppm 0.01 g/L |
| TDS Accuracy at 25 °C / 77 °F | ±1% of reading (±0.03 ppm or 1 digit, whichever is greater |
| Salinity* Range | 0.0 to 400.0 % NaCl 2.00 to 42.00 PSU 0.00 to 80.00 g/L |
| Salinity* Resolution | 0.1% NaCl 0.01 PSU 0.01 g/L |
| Salinity* Accuracy at 25 °C / 77 °F | ±1% of reading |
| Temperature Range (HI763100) | −20.0 to 120.0 °C**(−4.0 to 248.0 °F) |
| Temperature Resolution (HI763100) | 0.1 °C (0.1 °F) |
| Temperature Accuracy at 25 °C / 77 °F (HI763100) | ±0.5 °C (0.9 °F) |
| EC calibration | Cell factor calibration Six standards available: 84 µS/cm, 1413 μS/cm 5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm, 111.8 mS/cm One point offset: 0.00 μS/cm |
| "Salinity calibration* (% NaCl only)" | Single point with HI7037 standard |
| Temperature compensation | • Automatic, −5 to 100 °C (23 to 212 °F) • NoTC, can be selected to measure absolute conductivity |
| Conductivity temperature coefficient* | 0.00 to 6.00 %/°C (EC and TDS only) (1.90 %/°C default) |
| TDS factor* | 0.40 to 0.80 (0.50 default) |
| Logging | Up to 1000* (400) records organized in: • Manual log-on-demand, maximum 200 logs • Manual log-on-stability, maximum 200 logs • Interval logging*, maximum 600 logs (100 lots) |
| Notes | * When operating in Standard Mode only! ‡ Absolute conductivity (TDS) is the conductivity value without temperature compensation. |
| HI2600 - using the Dissolved Oxygen probe | HI764080 |
| DO Range | 0.00 to 45.00 ppm (mg/L) 0.0 to 300.0% |
| DO Resolution | 0.01 ppm (mg/L) 0.1% |
| DO Accuracy at 25 °C / 77 °F | ±1.5% of reading ±1 digit |
| Temperature Range | -20.0 to 120.0 °C * (−4.0 to 24 |
| Temperature Resolution | 0.1 °C (0.1 °F) |
| Temperature Accuracy at 25 °C / 77 °F | ±0.5 °C (0.9 °F) |
| DO calibration | One or two points at 0% (HI7040) and 100% (water-saturated air) |
| DO Pressure compensation | Automatic 450.0 to 850.0 mmHg |
| DO Pressure compensation resolution | 0.1 mmHg |
| DO Salinity compensation | 0 to 40 g/L |
| DO Salinity compensation resolution | 1 g/L |
| Temperature compensation | 0.0 to 50.0 °C (32.0 to 122 °F) |
| Logging | Up to 1000 records organized in: • Manual log-on-demand, maximum 200 logs • Manual log-on-stability, maximum 200 logs • Interval logging**, maximum 600 logs (100 lots) |
| Notes | * Temperature limits will be reduced to actual probe limits. ** When operating in Standard Mode only! |
| PC interface | USB-C |
| Power supply | USB Type C (5 VDC; 500 mA) |
| Environment | 0 to 50 °C (32 to 122 °F) Maximum 95 % RH non-condensing |
| Dimensions | 205 x 160 x 77 mm (8.0 x 6.2 x 3.0 inches) |
| Weight | Approximately 0.85 kg (1.87 lbs.) |