Functional and accurate, this meter is capable of testing 12 different water quality parameters using pH, EC, and optical DO sensors. Transfer data to a smart device for review or sharing with the integrated Bluetooth connection and Hanna Lab App.
*Products classified as Dangerous Goods can only be shipped via ground.
Measure up to 12 parameters. Don't lose important data with automatic interval logging of up to 45,000 samples or choose to log-on-demand to store measurement data for key points.
The meter operates on a built-in lithium-ion battery and will automatically switch to AA alkaline batteries for at least 200 hours of testing.
For better traceability, calibration information is recorded along with a time and date stamp and stored with logged data.
Quick Calibration provides a speedy, single point calibration for pH, conductivity, and dissolved oxygen. Standard calibration options are also available.
Designed to withstand the knocks, drops, and spills of real life, the IP67 body ensures top performance in any environment. These meters are protected against dust and water intrusion from any direction.
A Quick Connect DIN connector makes attaching and removing the probe simple and easy. The rubber coating protects the cable and provides a water-tight seal without twisting.
The USB Type-C port provides for easy data transfer to memory stick, PC, or other device and is used to charge internal lithium-ion battery.
Digital probe with built-in temperature sensor and three ports for pH (ORP), EC and optical DO sensors. Once connected, the meter aautomatically recognize the sensors.
Sensor replacement is quick and easy with field replaceable, screw type connectors and are color-coded for easy identification.
HI7698194-0 (pH only)
HI7698194-1 (pH/ORP)
HI7698194-3
HI7698494-5
* The optical dissolved oxygen sensor uses a smart cap that has an RFID tag that keeps track of the age of the cap and alerts you when it should be replaced.
Connect wirelessly via Bluetooth® to your smart device running the Hanna Lap App.
Select the measurement units to be displayed on your smart device. For trend analysis, take a look at the graph of your logged data.
Comprehensive GLP data can be reviewed for all parameters when the logged data is downloaded to your smart device.
Group your data by time or notation. Email it for storage or share it with friends or colleagues.
Try the Hanna Lab App in demo mode for free.
Specification | Detail |
---|---|
sku | HI98494-web |
pH Range | 0.00 to 14.00 pH |
pH Resolution | 0.01 pH |
pH Accuracy (@25°C/77°F) | ± 0.02 pH |
pH Calibration | automatic one, two, or three points with automatic recognition of five standard buffers (pH 4.01, 6.86, 7.01, 9.18, 10.01) or one custom buffer |
mv Range | ± 600.0 mV |
mv Resolution | 0.1 mV |
mv Accuracy | ± 0.5 mV |
ORP Range | ± 2000.0 mV |
ORP Resolution | 0.1 mV |
ORP Accuracy (@25°C/77°F) | ± 1.0 mV |
ORP Calibration | automatic at one custom point (relative mV) |
Temperature Range | -5.00 to 55.00 °C, 23.00 to 131.00 °F, 268.15 to 328.15 K |
Temperature Resolution | 0.01 °C, 0.01 °F, 0.01 K |
Temperature Accuracy (@25°C/77°F) | ± 0.15 °C, ± 0.27 °F, ±0.15 K |
Temperature Calibration | automatic at one custom point |
EC Range | 0 to 200 mS/cm (absolute EC up to 400 mS/cm) |
EC Resolution | Manual: 1 µS/cm, 0.001 mS/cm, 0.01 mS/cm, 0.1 mS/cm, 1 mS/cM Automatic: 1 µS/cm from 0 to 9999 µS/cm, 0.01 mS/cm from 10.00 to 99.99 mS/cm, 0.1 mS/cm from 100.0 to 400.0 mS/cM Automatic (mS/cm): 0.001 mS/cm from 0.000 to 9.999 |
EC Accuracy (@25°C/77°F) | ±1 % of reading or ±1 µS/cm whichever is greater |
EC Calibration | automatic single point, with six standard solutions (84 µS/cm, 1413 µS/cm, 5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm, 111.8 mS/cm) or custom point |
TDS Range | 0 to 400000 ppm (mg/L) |
TDS Resolution | Manual: 1 ppm (mg/L), 0.001 ppt (g/L), 0.01 ppt (g/L), 0.1 ppt (g/L), 1 ppt (g/L)Automatic: 1 ppm (mg/L) from 0 to 9999 ppm (mg/L), 0.01 ppt (g/L) from 10.00 to 99.99 ppt (g/L), 0.1 ppt (g/L) from 100.0 to 400.0 ppt (g/L), |
TDS Accuracy (@25°C/77°F) | ±1 % of reading or ±1 ppm (mg/L) whichever is greater |
TDS Calibration | based on conductivity or salinity calibration |
DO Range | 0.0 to 500.0 %, 0.00 to 50.00 ppm (mg/L) |
DO Resolution | 0.1 %, 0.01 ppm (mg/L) |
DO Accuracy (@25°C/77°F) | 0.0 to 300.0 %: ± 1.5 % of reading or ± 1.0 % whichever is greater, 300.0 to 500.0 %: ± 3 % of reading0.00 to 30.00 ppm (mg/L): ± 1.5 % of reading or ±0.10 ppm (mg/L)whichever is greater, 30.00 ppm (mg/L) to 50.00 ppm (mg/L): ± 3 % of reading |
DO Calibration | automatic one or two points at 0 and 100% or one custom point |
Resistivity Range | 0 to 999999 ©·cm, 0 to 1000.0 k©·cm, 0 to 1.0000 M©·cm |
Resistivity Resolution | Depending on resistivity reading |
Resistivity Accuracy (@25°C/77°F) | Based on conductivity or salinity calibration |
Salinity Range | 0.00 to 70.00 PSU |
Salinity Resolution | 0.01 PSU |
Salinity Accuracy (@25°C/77°F) | ±2% of reading or ±0.01 PSU whichever is greater |
Salinity Calibration | based on conductivity calibration |
Connectivity | USB to PC with HI9298194 Hanna software installed |
GLP | Yes |
Power Supply | 1.5V AA batteries (4) |
Power Consumption | approximately 360 hours of continuous use without backlight (50 hours with backlight) |
Environment | 0 to 50°C (32 to 122°F), RH 100% (IP67) |
Dimensions | 185 x 93 x 35.2 mm (7.3 x 3.6 x 1.4in.) |
Weight | 400 g (14.2 oz.) |
Order Information | Each HI98494 is delivered in a rugged carrying case and is supplied with: HI7698494 Multi-sensor probe, HI7698194-1 pH/ORP sensor, HI7698194-3 EC/TDS/Resistivity/Salinity sensor, HI7698494-5 Optical DO sensor, HI764113-1 DO Smart Cap |
Warranty | 2 years |
Recommended Probe | HI7698494 multiparameter probe, HI7698194-1 pH sensor, HI7698194-1 pH/ORP sensor, HI7698194-3 EC sensor, HI7698494-5 optical DO sensor |