Our Disinfectants Multiparameter Portable Photometer (HI97107) measures seven important parameters in the treatment and disinfection of drinking water and wastewater.
Our Portable Photometers benefits from Hanna’s® years of experience as a manufacturer of analytical instruments. It has an advanced optical system that uses a Light Emitting Diode (LED) and a narrow band interference filter that allows for accurate and repeatable readings.
*Products classified as Dangerous Goods can only be shipped via ground.
The optical system is sealed from outside dust, dirt and water. The meter uses an exclusive positive-locking system to ensure that the cuvettes are placed into the holder in the same position every time.
With the CAL Check™ functionality, users are able to validate the performance of the instrument at any time and apply a user calibration (if necessary). Hanna Instruments CAL Check™ cuvettes are made with NIST traceable standards.
The built-in tutorial mode guides users step-by-step through the measurement process. It includes all steps required for sample preparation, the required reagents and quantities.
Bromine • Hydrogen Peroxide • Ozone • pH • Free & Total Chlorine LR • Total Chlorine UHR • Chlorine Dioxide (Rapid)
Bromine is a powerful water disinfectant, particularly effective in hot water (treatment of pools and spas). Bromine can act as an alternative to chlorine.
The method for bromine is an adaptation of Standard Methods for the Examination of Water and Wastewater, DPD Method.
Chlorine is a widely used disinfectant, in order for it to be effective the pH of the water should be less than 8.0.
The method for chlorine is an adaptation of US EPA Method 330.5, DPD Colorimetric Method. The method for pH is an adaptation of the Phenol Red Method.
is a powerful and selective disinfectant used in water treatment across a broad pH range.
The method for chlorine dioxide is an adaptation of Standard Methods for the Examination of Water and Wastewater, 18th Ed., 4500 ClO₂ D.
Hydrogen Peroxide is an environmentally friendly, broad spectrum disinfectant.
The method for hydrogen peroxide is an adaptation of German Standard Methods for the Examination of Water and Wastewater DIN 38409-15.
Ozone is a widely used disinfectant in the treatment of drinking water, wastewater and swimming pools.
The method for ozone is the DPD Colorimetric Method.
Let your photometer guide you through setup, calibration, and testing with on-screen tutorials and animations so you always know the next step. Simply access Tutorial Mode at any stage in the setup or measurement process with the dedicated help key.
The auto-off feature prolongs battery life for peace-of-mind and energy savings if the meter is accidentally left on.
When the battery level is below 10%, a warning symbol will blink to indicate low battery condition. When low battery power may affect your results, the meter will shut down to avoid erroneous readings caused by low battery level.
Missed a measurement? No need to worry because your photometer will automatically log the last 50 measurements. The results can be viewed instantly using the Log Recall Menu.
Hanna's exclusive CAL Check feature provides a stress-free, straightforward way to ensure your photometer is working properly. Simply insert the certified CAL Check standards and let your meter guide you through validation. If your photometer is out of specification for any reason, simply calibrate to your standard.
A Light Emitting Diode (LED) provides more light while using less power and a narrow band interference filter results in more accurate and repeatable readings.
Designed to withstand the knocks, drops, and spills of real life, the new IP67 body ensures top performance in any environment. Internal optics are sealed from outside dust, dirt, and water. In addition, it floats, just in case you drop your meter in water. This photometer uses an exclusive positive?locking system to ensure that the cuvettes are placed into the holder in the same position for accurate readings every time.
Made to fit comfortably in your hand for on-the-go testing or sit on a table for benchtop use, a backlit LCD display provides at-a-glance readings from any angle, day or night. The supplied AA batteries provide life for more than 800 measurements.
Each HI97107 is delivered in a cardboard box and is supplied with:
Each HI97107C is delivered in a rugged carrying case and is supplied with:
| Specification | Detail |
|---|---|
| Bromine Range | 0.00 to 10.00 mg/L (as Br2) |
| Bromine Resolution | 0.01 mg/L |
| Bromine Accuracy | ±0.08 mg/L ±3% of reading at 25 °C (77 °F) |
| Bromine Method | Adaptation of Standard Methods for the Examination of Water and Wastewater, 18th Edition, DPD Method |
| Free and Total Chlorine LR Range | 0.00 to 5.00 mg/L (as Cl2 |
| Free and Total Chlorine LR Resolution | 0.01 mg/L |
| Free and Total Chlorine LR Accuracy | ±0.03 mg/L ±3% of reading at 25 °C (77 °F) |
| Free and Total Chlorine LR Method | Adaptation of US EPA Method 330.5, DPD Colorimetric Method |
| Total Chlorine UHR Range | 0 to 500 mg/L (as Cl2) |
| Total Chlorine UHR Resolution | 1 mg/L |
| Total Chlorine UHR Accuracy | ±3 mg/L ±3 % of reading at 25 °C (77 °F) |
| Total Chlorine UHR Method | Adaptation of Standard Methods for Examination of Water and Wastewater, |
| Chlorine Dioxide (Rapid) Range | 0.00 to 2.00 mg/L (as ClO2) |
| Chlorine Dioxide (Rapid) Resolution | 0.01 mg/L |
| Chlorine Dioxide (Rapid) Accuracy | ±0.10 mg/L ±5% of reading at 25 °C (77 °F) |
| Chlorine Dioxide (Rapid) Method | Adaptation of Standard Methods for the Examination of Water and Wastewater, 18th Edition, 4500 ClO2 D |
| Hydrogen Peroxide Range | 0 to 500 mg/L H2O2 |
| Hydrogen Peroxide Resolution | 1 mg/L |
| Hydrogen Peroxide Accuracy | ±5 mg/L ±3 % of reading at 25 °C (77 °F) |
| Hydrogen Peroxide Method | Adaptation of DIN 38409-15, German Standard Methods for the Examination of Water and Wastewater |
| Ozone Range | 0.00 to 2.00 mg/L (as O3) |
| Ozone Resolution | 0.01 mg/L |
| Ozone Accuracy | ±0.02 mg/L ±3 % of reading at 25 °C (77 °F) |
| Ozone Method | DPD Colorimetric Method |
| pH Range | 6.5 to 8.5 pH |
| pH Resolution | 0.1 pH |
| pH Accuracy | ±0.1 pH of reading at 25 °C (77 °F) |
| pH Method | Adaptation of the Phenol Red Method |
| Light source | Light Emitting Diode |
| Bandpass filter | 525 nm |
| Bandpass filter bandwidth | 8 nm |
| Bandpass filter wavelength accuracy | ±1.0 nm |
| Light detector | Silicon photocell |
| Cuvette type | Round 24.6 mm diameter (22 mm inside) |
| Auto logging | 50 readings |
| Display | 128×64 pixel B/W LCD with backlight |
| Auto-off | After 15 minutes of inactivity (30 minutes before a READ measurement) |
| Battery type | 1.5 V AA Alkaline (3 pcs.) |
| Battery life | >800 measurements (without backlight) |
| Environment | 0 to 50 °C (32 to 122 °F) 0 to 100% RH, non-serviceable |
| Dimensions | 142.5×102.5×50.5 mm (5.6×4.0×2.0 in.) |
| Weight (with batteries) | 380 g (13.4 oz.) |
| Case ingress protection rating | IP67, floating case |